Building Block Commands

Building block, or basic, commands are the smallest individual actions that can be completed on an OT-2. For example, the complex command transfer (see Complex Commands) executes a series of pick_up_tip(), aspirate(), dispense() and drop_tip() basic commands.

The examples in this section would be added to the following:

from opentrons import protocol_api

metadata = {'apiLevel': '2.11'}

def run(protocol: protocol_api.ProtocolContext):
    tiprack = protocol.load_labware('corning_96_wellplate_360ul_flat', 2)
    plate = protocol.load_labware('opentrons_96_tiprack_300ul', 3)
    pipette = protocol.load_instrument('p300_single_gen2', mount='left')
    # the example code below would go here, inside the run function

This loads a Corning 96 Well Plate in slot 2 and a Opentrons 300 µL Tiprack in slot 3, and uses a P300 Single GEN2 pipette.

Tip Handling

When the OT-2 handle liquids with, it constantly exchanges old, used tips for new ones to prevent cross-contamination between wells. Tip handling uses the functions InstrumentContext.pick_up_tip(), InstrumentContext.drop_tip(), and InstrumentContext.return_tip().

Pick Up Tip

Before any liquid handling can be done, your pipette must have a tip on it. The command InstrumentContext.pick_up_tip() will move the pipette over to the specified tip, then press down into it to create a vacuum seal. The below example picks up the tip at location 'A1' of the tiprack previously loaded in slot 3.

pipette.pick_up_tip(tiprack['A1'])

If you have associated a tiprack with your pipette such as in the Pipettes or Protocols and Instruments sections, then you can simply call

pipette.pick_up_tip()

This will use the next available tip from the list of tipracks passed in to the tip_racks argument of ProtocolContext.load_instrument().

New in version 2.0.

Drop Tip

Once finished with a tip, the pipette will remove the tip when we call InstrumentContext.drop_tip(). You can specify where to drop the tip by passing in a location. The below example drops the tip back at its original location on the tip rack. If no location is specified, the OT-2 will drop the tip in the fixed trash in slot 12 of the deck.

pipette.pick_up_tip()
pipette.drop_tip(tiprack['A1'])  # drop back in A1 of the tiprack
pipette.pick_up_tip()
pipette.drop_tip()  # drop in the fixed trash on the deck

New in version 2.0.

Return Tip

To return the tip to the original location, you can call InstrumentContext.return_tip(). The example below will automatically return the tip to 'A3' on the tip rack.

pipette.pick_up_tip(tiprack['A3'])
pipette.return_tip()

In API version 2.2 or above:

tip_rack = protocol.load_labware(
        'opentrons_96_tiprack_300ul', 1)
pipette = protocol.load_instrument(
    'p300_single_gen2', mount='left', tip_racks=[tip_rack])

pipette.pick_up_tip() # picks up tip_rack:A1
pipette.return_tip()
pipette.pick_up_tip() # picks up tip_rack:B1

In API version 2.0 and 2.1:

tip_rack = protocol.load_labware(
        'opentrons_96_tiprack_300ul', 1)
pipette = protocol.load_instrument(
    'p300_single_gen2', mount='left', tip_racks=[tip_rack])

pipette.pick_up_tip() # picks up tip_rack:A1
pipette.return_tip()
pipette.pick_up_tip() # picks up tip_rack:A1

Iterating Through Tips

For this section, instead of using the protocol defined above, consider this setup:

from opentrons import protocol_api

metadata = {'apiLevel': '2.11'}

def run(protocol: protocol_api.ProtocolContext):
    plate = protocol.load_labware(
        'corning_96_wellplate_360ul_flat', 2)
    tip_rack_1 = protocol.load_labware(
        'opentrons_96_tiprack_300ul', 3)
    tip_rack_2 = protocol.load_labware(
        'opentrons_96_tiprack_300ul', 4)
    pipette = protocol.load_instrument(
        'p300_single_gen2', mount='left', tip_racks=[tip_rack_1, tip_rack_2])

This loads a Corning 96 Well Plate in slot 2 and two Opentrons 300ul Tiprack in slots 3 and 4 respectively, and uses a P300 Single GEN2 pipette.

When a list of tip racks is associated with a pipette in its tip_racks argument, the pipette will automatically pick up the next unused tip in the list whenever you call InstrumentContext.pick_up_tip(). The pipette will first use all tips in the first tiprack, then move on to the second, and so on:

pipette.pick_up_tip()  # picks up tip_rack_1:A1
pipette.return_tip()
pipette.pick_up_tip()  # picks up tip_rack_1:A2
pipette.drop_tip()     # automatically drops in trash

# use loop to pick up tips tip_rack_1:A3 through tip_rack_2:H12
tips_left = 94 + 96 # add up the number of tips leftover in both tipracks
for _ in range(tips_left):
    pipette.pick_up_tip()
    pipette.return_tip()

If you try to InstrumentContext.pick_up_tip() again when all the tips have been used, the Protocol API will show you an error:

# this will raise an exception if run after the previous code block
pipette.pick_up_tip()

To change the location of the first tip used by the pipette, you can use InstrumentContext.starting_tip:

pipette.starting_tip = tip_rack_1.well('C3')
pipette.pick_up_tip()  # pick up C3 from "tip_rack_1"
pipette.return_tip()

To reset the tip tracking, you can call InstrumentContext.reset_tipracks():

# Use up all tips
for _ in range(96+96):
     pipette.pick_up_tip()
     pipette.return_tip()

# Reset the tip tracker
pipette.reset_tipracks()

# Picks up a tip from well A1 of the first tip rack
pipette.pick_up_tip()

New in version 2.0.

To check whether you should pick up a tip or not, you can utilize InstrumentContext.has_tip():

for block in range(3):
    if block == 0 and not pipette.has_tip:
        pipette.pick_up_tip()
    else:
        m300.mix(mix_repetitions, 250, d)
        m300.blow_out(s.bottom(10))
        m300.return_tip()

New in version 2.7.


Liquid Control

This section describes the InstrumentContext ‘s liquid-handling commands.

The examples in this section should be inserted in the following:

metadata = {'apiLevel': '2.11'}

def run(protocol):
    plate = protocol.load_labware('corning_96_wellplate_360ul_flat', 2)
    tiprack = protocol.load_labware('opentrons_96_tiprack_300ul', 3)
    pipette = protocol.load_instrument('p300_single_gen2', mount='left', tip_racks=[tiprack])
    pipette.pick_up_tip()
    # example code goes here

This loads a Corning 96 Well Plate in slot 2 and a Opentrons 300ul Tiprack in slot 3, and uses a P300 Single GEN2 pipette.

Aspirate

To aspirate is to pull liquid up into the pipette’s tip. When calling InstrumentContext.aspirate() on a pipette, you can specify the volume to aspirate in µL, where to aspirate from, and how fast to aspirate liquid.

pipette.aspirate(50, plate['A1'], rate=2.0)  # aspirate 50uL from plate:A1

Now the pipette’s tip is holding 50 µL.

The location parameter is either a well (like plate['A1']) or a position within a well, like the return value of plate['A1'].bottom.

The rate parameter is a multiplication factor of the pipette’s default aspiration flow rate. The default aspiration flow rate for all pipettes is in the Defaults section.

You can also simply specify the volume to aspirate, and not mention a location. The pipette will aspirate from its current location (which we previously set as plate['A1']).

pipette.aspirate(50)                     # aspirate 50uL from current position

Now our pipette’s tip is holding 100 µL.

Note

In version 1 of this API, aspirate (and dispense) would inspect the types of the volume and location arguments and do the right thing if you specified only a location or specified location and volume out of order. In this and future versions of the Python Protocol API, this is no longer true. Like any other Python function, if you are specifying arguments by position without using their names, you must always specify them in order.

Note

By default, the OT-2 will move to 1mm above the bottom of the target well before aspirating. You can change this by using a well position function like Well.bottom() (see Specifying Position Within Wells) every time you call aspirate, or - if you want to change the default throughout your protocol - you can change the default offset with InstrumentContext.well_bottom_clearance (see Default Positions Within Wells).

New in version 2.0.

Dispense

To dispense is to push out liquid from the pipette’s tip. The usage of InstrumentContext.dispense() in the Protocol API is similar to InstrumentContext.aspirate(), in that you can specify volume in µL and location, or only volume.

pipette.dispense(50, plate['B1'], rate=2.0) # dispense 50uL to plate:B1 at twice the normal rate
pipette.dispense(50)              # dispense 50uL to current position at the normal rate

The location parameter is either a well (like plate['A1']) or a position within a well, like the return value of plate['A1'].bottom.

The rate parameter is a multiplication factor of the pipette’s default dispense flow rate. The default dispense flow rate for all pipettes is in the Defaults section.

Note

By default, the OT-2 will move to 1mm above the bottom of the target well before dispensing. You can change this by using a well position function like Well.bottom() (see Specifying Position Within Wells) every time you call dispense, or - if you want to change the default throughout your protocol - you can change the default offset with InstrumentContext.well_bottom_clearance (see Default Positions Within Wells).

Note

In version 1 of this API, dispense (and aspirate) would inspect the types of the volume and location arguments and do the right thing if you specified only a location or specified location and volume out of order. In this and future versions of the Python Protocol API, this is no longer true. Like any other Python function, if you are specifying arguments by position without using their names, you must always specify them in order.

New in version 2.0.

Blow Out

To blow out is to push an extra amount of air through the pipette’s tip, to make sure that any remaining droplets are expelled.

When calling InstrumentContext.blow_out(), you can specify a location to blow out the remaining liquid. If no location is specified, the pipette will blow out from its current position.

pipette.blow_out()            # blow out in current location
pipette.blow_out(plate['B3']) # blow out in current plate:B3

New in version 2.0.

Touch Tip

To touch tip is to move the pipette’s currently attached tip to four opposite edges of a well, to knock off any droplets that might be hanging from the tip.

When calling InstrumentContext.touch_tip() on a pipette, you have the option to specify a location where the tip will touch the inner walls.

InstrumentContext.touch_tip() can take up to 4 arguments: touch_tip(location, radius, v_offset, speed).

pipette.touch_tip()            # touch tip within current location
pipette.touch_tip(v_offset=-2) # touch tip 2mm below the top of the current location
pipette.touch_tip(plate['B1']) # touch tip within plate:B1
pipette.touch_tip(plate['B1'], speed=100) # touch tip within plate:B1 at 100 mm/s
pipette.touch_tip(plate['B1'], # touch tip in plate:B1, at 75% of total radius and -2mm from top of well
                  radius=0.75,
                  v_offset=-2)

New in version 2.0.

Mix

To mix is to perform a series of aspirate and dispense commands in a row on a single location. Instead of having to write those commands out every time, you can call InstrumentContext.mix().

The mix command takes up to three arguments: mix(repetitions, volume, location):

# mix 4 times, 100uL, in plate:A2
pipette.mix(4, 100, plate.['A2'])
# mix 3 times, 50uL, in current location
pipette.mix(3, 50)
# mix 2 times, pipette's max volume, in current location
pipette.mix(2)

Note

In API Versions 2.2 and earlier, mixes consist of aspirates and then immediate dispenses. In between these actions, the pipette moves up and out of the target well. In API Version 2.3 and later, the pipette will not move between actions.

New in version 2.0.

Air Gap

When dealing with certain liquids, you may need to aspirate air after aspirating the liquid to prevent it from sliding out of the pipette’s tip. A call to InstrumentContext.air_gap() with a volume in µL will aspirate that much air into the tip. air_gap takes up to two arguments: air_gap(volume, height):

pipette.aspirate(100, plate['B4'])
pipette.air_gap(20)
pipette.drop_tip()

New in version 2.0.


Utility Commands

Move To

You can use InstrumentContext.move_to() to move a pipette to any location on the deck.

For example, you can move to the first tip in your tip rack:

pipette.move_to(tiprack['A1'].top())

Unlike commands that require labware, like Aspirate or Dispense, InstrumentContext.move_to() deals with types.Location instances, which combine positions in Deck Coordinates and associated Labware instances. You don’t have to create them yourself; this is what is returned from methods such as Well.top() and Well.bottom(). It does mean, however, that you can’t move to a well directly; you must use Well.top() or build a types.Location yourself.

You can also specify at what height you would like the robot to move to inside of a location using Well.top() and Well.bottom() methods on that location (more on these methods and others like them in the Specifying Position Within Wells section):

pipette.move_to(plate['A1'].bottom())  # move to the bottom of well A1
pipette.move_to(plate['A1'].top())     # move to the top of well A1
pipette.move_to(plate['A1'].bottom(2)) # move to 2mm above the bottom of well A1
pipette.move_to(plate['A1'].top(-2))   # move to 2mm below the top of well A1

The above commands will cause the robot’s head to first move upwards, then over to above the target location, then finally downwards until the target location is reached. If instead you would like the robot to move in a straight line to the target location, you can set the movement strategy to 'direct'.

pipette.move_to(plate['A1'].top(), force_direct=True)

Warning

Moving without an arc will run the risk of colliding with things on your deck. Be very careful when using this option.

Usually the above option is useful when moving inside of a well. Take a look at the below sequence of movements, which first move the head to a well, and use ‘direct’ movements inside that well, then finally move on to a different well.

pipette.move_to(plate['A1'].top())
pipette.move_to(plate['A1'].bottom(1), force_direct=True)
pipette.move_to(plate['A1'].top(-2), force_direct=True)
pipette.move_to(plate['A2'].top())

New in version 2.0.

Delay for an Amount of Time

Sometimes you need to wait as a step in your protocol, for instance to wait for something to incubate. You can use ProtocolContext.delay() to wait your protocol for a specific amount of time. delay is a method of ProtocolContext since it concerns the protocol and the OT-2 as a whole.

The values passed into delay() specify the number of minutes and seconds that the OT-2 will wait until moving on to the next command.

protocol.delay(seconds=2)             # delay for 2 seconds
protocol.delay(minutes=5)             # delay for 5 minutes
protocol.delay(minutes=5, seconds=2)  # delay for 5 minutes and 2 seconds

Pause Until Resumed

The method ProtocolContext.pause() will pause protocol execution at a specific step. You can resume by pressing ‘resume’ in your Opentrons App. You can optionally specify a message that will be displayed in the Opentrons App when protocol execution pauses.

from opentrons import protocol_api

metadata = {'apiLevel': '2.11'}

def run(protocol: protocol_api.ProtocolContext):
    # The start of your protocol goes here...

    # The OT-2 stops here until you press resume. It will display the message in
    # the Opentrons App. You do not need to specify a message, but it makes things
    # more clear.
    protocol.pause('Time to take a break')

New in version 2.0.

Homing

You can manually request that the OT-2 home during protocol execution. This is typically not necessary; however, if at any point you will disengage motors or move the gantry with your hand, you may want to command a home afterwards.

To home the entire OT-2, you can call ProtocolContext.home().

To home a specific pipette’s Z axis and plunger, you can call InstrumentContext.home().

To home a specific pipette’s plunger only, you can call InstrumentContext.home_plunger().

None of these functions take any arguments:

from opentrons import protocol_api, types

metadata = {'apiLevel': '2.11'}

def run(protocol: protocol_api.ProtocolContext):
    pipette = protocol.load_instrument('p300_single', 'right')
    protocol.home() # Homes the gantry, z axes, and plungers
    pipette.home()  # Homes the right z axis and plunger
    pipette.home_plunger() # Homes the right plunger

New in version 2.0.

Comment

The method ProtocolContext.comment() lets you display messages in the Opentrons App during protocol execution:

from opentrons import protocol_api, types

metadata = {'apiLevel': '2.11'}

def run(protocol: protocol_api.ProtocolContext):
    protocol.comment('Hello, world!')

New in version 2.0.

Control and Monitor Robot Rail Lights

You can turn the robot rail lights on or off in the protocol using ProtocolContext.set_rail_lights():

from opentrons import protocol_api

metadata = {'apiLevel': '2.11'}

def run(protocol: protocol_api.ProtocolContext):
    # turn on robot rail lights
    protocol.set_rail_lights(True)

    # turn off robot rail lights
    protocol.set_rail_lights(False)

New in version 2.5.

You can also check whether the rail lights are on or off in the protocol using ProtocolContext.rail_lights_on:

protocol.rail_lights_on  # returns True when the lights are on,
                         # False when the lights are off

New in version 2.5.

Monitor Robot Door

The door safety switch feature flag has been added to the OT-2 software since the 3.19.0 release. Enabling the feature flag allows your robot to pause a running protocol and prohibit the protocol from running when the robot door is open.

_images/door_safety_switch.png

You can also check whether or not the robot door is closed at a specific point in time in the protocol using ProtocolContext.door_closed:

protocol.door_closed  # return True when the door is closed,
                      # False when the door is open

Note

Both the top window and the front door must be closed in order for the robot to report the door is closed.

Warning

If you chose to enable the door safety switch feature flag, you should only use ProtocolContext.door_closed as a form of status check, and should not use it to control robot behavior. If you wish to implement custom method to pause or resume protocol using ProtocolContext.door_closed, make sure you have first disabled the feature flag.

New in version 2.5.